Product Cat. No.: GBS-107
For Research Use Only.
Chromosome 3 centromere probe reagent
10 Tests/box
This kit uses orange fluorescein or green fluorescein to mark the CEP3 probe. The CEP3 probe can be combined with the target detection site by in situ hybridization technology.
The kit offers two types of specifications respectively. Spec 1 consists of CEP3 green probe, Spec 2 consists of CEP3 orange probe, as shown in Table 1.
Component name | Specifications | Quantity | Main components |
---|---|---|---|
CEP3 Green probe | 100μL/Tube | 1 | CEP3 Green probe |
CEP3 Orange probe | 100μL/Tube | 1 | CEP3 Orange probe |
Keep sealed away from light at -20oC±5oC. The product is valid for 12 months. Avoid unnecessary repeated freezing and thawing that should not exceed 10 times. After opening, within 24 hours for short-term preservation, keep sealed at 2-8oC in dark. For long-term preservation after opening, keep the lid sealed at -20oC±5oC away from light. The kit is transported below 0oC.
Fluorescence microscopy imaging systems, including fluorescence microscopy and filter sets suitable for DAPI (367/452), Green (495/517),and Orange (547/565).
Tissue samples:
1. Applicable specimen type: paraffin embedded specimen of surgical resection or biopsy tissue.
2. The tissue should be fixed with 4% neutral formaldehyde fixation solution. After tissue fixation, it should be regularly dehydrated and embedded in paraffin.
Cell sample:
1. Applicable specimen type: unfixed fresh bone marrow specimen (stored at 2-8 °C for no more than 24 hours).
2. Sample collection take 1-3ml of bone marrow cells anticoagulated with heparin sodium.
3. Sample preservation: after fixation, the cell suspension shall be stored at – 20 ± 5 °C for no more than 12 months; The prepared cell slides can be stored at – 20 ± 5 °C for no more than 1 month. When the sample storage temperature is too high or too low, or the cell suspension is volatilized excessively or polluted during storage, the sample will not be used for detection.
The following reagents are required for the experiment but not provided in this kit.
1. Sample collection and slides preparation
Tissue samples:
Baking: Slides heating at 80oC for 30min or 65oC for 2h or overnight.
Dewaxing: According to the customer laboratory protocol (Commonly with Xylene for 15min).
Hydration: Take out the slides and put them respectively into 100%, 85% and 70% EtOH at room temperature for 3 minutes each.
Take out the slides, and immerse them in deionized water for 3 minutes. Remove the excess of water on the slides by air-drying.
Permeation: Immerse the slides in deionized water at 100oC and boil continuously for 20-40 minutes (Conventional 20min). Remove the excess of water on the slides by air-drying.
Digestion: Protease enzymic digestion at 37°C for 10-40 minutes. Mix the protease work buffer (50mmol HCl) and the 10x protease solution (Pepsin concentration 0.5%) in a proportion of 9:1 to prepare the enzymatic digestion solution.
Washing: Wash with 2xSSC at room temperature for 5 minutes.
Dehydration: Take out the slides and dehydrate in 70%, 85%, and 100% gradient ethanol at room temperature for 2 minutes each time.Remove the excess of EtOH solution on the slides by air-drying.
Cell sample:
2. Denaturation and Hybridization
The following operations should be performed in a darkroom.
Tissue samples:
Cell sample:
3. Washing
The following operations should be performed in a darkroom.
4. Counterstaining
The following operations should be performed in a darkroom
10μL DAPI compound dye is dropped in the hybridization area of the glass slide and immediately covered. The suitable filter is selected for glass slide observation under the fluorescence microscope.
5. FISH results observation
Put the counterstained cell drops under the fluorescence microscope, first under the low-power objective lens (10x) confirm the cell area under the microscope; Go to 40x under the objective lens, find a position where the cells are evenly distributed; Then in the high-power objective (100x) the FISH results of nuclei were observed.
1.The results of this kit will be affected by various factors of the sample itself, but also limited by hybridization temperature and time, operating environment and the limitations of current molecular biology technology, which may lead to wrong results.
2.Users must understand the potential errors and accuracy limitations that may exist in the detection process.
3.All chemicals are potentially dangerous. Avoid direct contact. Used kits are clinical waste and should be properly disposed off.
At Genebiosolution, our commitment lies in crafting cutting-edge solutions for researchers, clinicians, and diagnostic laboratories seeking excellence in immunohistochemistry (IHC) and fluorescence in situ hybridization (FISH). With a focus on quality, reliability, and accuracy, our products empower advancements in molecular diagnostics and research methodologies.